| Literature DB >> 6798394 |
N B Tannous, W F Gagnon, P R Almond.
Abstract
Buildup and surface-dose measurements were taken for the 6 MV photon beam from a Therac 6 linear accelerator manufactured by Atomic Energy of Canada Limited (AECL) with and without a lucite blocking tray in place. Further measurements were made with a copper filter designed to reduce secondary electrons emitted by photon interactions with the Lucite tray. The results are discussed in relation to skin-sparing for radiation therapy patients. The measurements were made with a fixed volume PTW parallel-plate ionization chamber and corrected to zero-chamber volume. The results were found to be consistent with similar measurements taken with a variable volume extrapolation chamber.Entities:
Mesh:
Year: 1981 PMID: 6798394 DOI: 10.1118/1.594961
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071