Literature DB >> 6794940

Comparison of gas-liquid chromatography and EMIT assay for serum valproic acid.

M Leroux, D Budnik, K Hall, J Irvine-Meek, N Otten, S Seshia.   

Abstract

We describe a simple direct extraction method for the gas-liquid chromatography determination of serum valproic acid. The working range for the assay is 2-180 mg/L and our within-run precision was 5.8 and 4.3% at the 40 and 90 mg/L concentrations respectively. Hemolyzed and lipemic sera as well as samples from patients with hyperbilirubinemia and from patients with decreased renal function were put through the assay and no interfering peaks were noted. Interference occurred when teflon-lined screw caps were used during the extraction step. The method was proven to be accurate by linear regression analysis of samples containing weighed-in amounts of valproic acid. The above assay was compared to an enzyme immunoassay technique (EMIT). The working range for the latter is 10-150 mg/L and the with-run precision was 10.8 and 5.9% and 90 mg/L concentration respectively. Samples were run by both the gas-liquid chromatograph and enzyme immunoassay methods and gave very similar results over the range 16-139 mg/L.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 6794940     DOI: 10.1016/s0009-9120(81)90772-4

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  1 in total

1.  First-dose and steady-state pharmacokinetics of valproic acid in children with seizures.

Authors:  K Hall; N Otten; J Irvine-Meek; M Leroux; D Budnick; M Verma; S S Seshia
Journal:  Clin Pharmacokinet       Date:  1983 Sep-Oct       Impact factor: 6.447

  1 in total