Bacterial contamination of anaesthetic gases.

H Nielsen, M Vasegaard, D B Stokke.   

Abstract

The bacterial content of oxygen and nitrous oxide immediately before and after passing through clean and used breathing systems (circuits) was measured using a specially constructed agar chamber (Bourdillon's slit sampler). The content per litre of oxygen from the outlet of the anaesthetic machine was 4.0 X 10-2, and 2.9 X 10-2 for nitrous oxide, corresponding to 3.5 X 10-2 for a 50% mixture of the gases. After passing through cleaned circuits, the bacterial pollution of the gas mixture had increased by 30%, but more than elevenfold after passing through used circuits. The content from cleaned circuits was less than that measured previously in the air of hospital wards and operating theatres, whereas gases from used circuits were polluted to approximately the same extent. It is concluded that used circuits may increase the risk of cross-infection. The cleaning method employed by us (dish-washer--hot airy drying) appeared to be acceptable.