| Literature DB >> 6766247 |
A H Barrett, P C Myers, N L Sadowsky.
Abstract
Microwave thermography, a method of sensing subcutaneous temperatures, was used in a breast cancer detection study of about 5,000 female patients. The data were taken at wavelengths of 9.1 and 23 cm. Microwave thermography at 23 cm has true-positive and true-negative detection rates of 0.8 and 0.6, respectively, comparable to those of infrared thermography (0.7) and inferior to those of xeromammography (0.9). However, a potential advantage results if microwave and infrared thermography are used together for screening, and if mammography is used only for follow-up on those patients who were positive on either the microwave or the infrared thermograms. It is then possible to obtain true-positive and true-negative detection rates of 0.9 and 0.9, respectively, while only half the number of patients need be subjected to x-rays.Entities:
Mesh:
Year: 1980 PMID: 6766247 DOI: 10.2214/ajr.134.2.365
Source DB: PubMed Journal: AJR Am J Roentgenol ISSN: 0361-803X Impact factor: 3.959