Clinical evaluation of the Dinamap non-invasive blood pressure monitor in pre-term neonates.

H D Dellagrammaticas1, A J Wilson.   

Abstract

The performance of the Dinamap 847 non-invasive blood pressure monitor was evaluated in a group of 10 severely ill pre-term newborn babies. The mean gestational age was 30 weeks and mean birthweight of 1610 gm. The values of systolic, diastolic and mean arterial pressure obtained by the Dinamap were compared with the corresponding values of actual pressure through a right radial artery catheter. All these values showed a high degree of correlation. No complications were noted. It was concluded that the Dinamap 847 should be considered as an alternative for monitoring the arterial blood pressure in pre-term neonates where arterial catheterisation is either not justified or impossible.

Mesh:

Year:  1981        PMID: 6765811     DOI: 10.1088/0143-0815/2/4/003

Source DB:  PubMed          Journal:  Clin Phys Physiol Meas        ISSN: 0143-0815


  4 in total

1.  Blood pressure monitoring: automated oscillometric devices.

Authors:  M Ramsey
Journal:  J Clin Monit       Date:  1991-01

2.  Systolic blood pressure in babies of less than 32 weeks gestation in the first year of life. Northern Neonatal Nursing Initiative.

Authors: 
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-01       Impact factor: 5.747

3.  Dinamap fails to detect hypotension in very low birthweight infants.

Authors:  G K Diprose; D H Evans; L N Archer; M I Levene
Journal:  Arch Dis Child       Date:  1986-08       Impact factor: 3.791

Review 4.  Hypotension in the very low birthweight infant: the old, the new, and the uncertain.

Authors:  S J Dasgupta; A B Gill
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-11       Impact factor: 5.747

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.