Corrosion of aneurysm clips: evaluation and clinical implications. Part 1: Metallurgical foundations of stress corrosion failure.

R Kossowsky, M Dujovny, N Kossovsky, F G Diaz, J I Ausman.   

Abstract

Several intracranial failures of aneurysm clips have been documented recently. Consistently, the mechanism of failure has been stress corrosion. These failures could have been anticipated if certain metallurgical properties of the clips had been known. In this paper, we review the metallurgical basis of stress corrosion failure and offer serial metallurgical tests as screening procedures to determine a clip's propensity towards failure. The importance of determining a clip's fracture toughness, a key indicator of stress corrosion resistance, is emphasized.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6741646     DOI: 10.1007/BF01406812

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  7 in total

1.  [Unusual recurrence of an IC--PC aneurysm after neck clipping--any structural weakness of the clip? (author's transl)].

Authors:  I Hayakawa; A Sasaki; K Fujiwara; T Tsuchida
Journal:  Acta Neurochir (Wien)       Date:  1976       Impact factor: 2.216

2.  Delayed breaking of a Heifetz aneurysm clip. Case report.

Authors:  A Servo; M Puranen
Journal:  J Neurosurg       Date:  1977-09       Impact factor: 5.115

3.  Fractured aneurysm clip. Case report.

Authors:  D O Quest; R W Countee
Journal:  J Neurosurg       Date:  1977-09       Impact factor: 5.115

4.  Failure of a Heifetz aneurysm clip.

Authors:  R Kossowsky; M Dujovny; N Kossovsky; Y Keravel
Journal:  J Neurosurg       Date:  1982-08       Impact factor: 5.115

5.  Metallurgical evaluation of the compatibility of surgical clips with their appliers.

Authors:  R Kossowsky; M Dujovny; N Kossovsky
Journal:  Acta Neurochir (Wien)       Date:  1981       Impact factor: 2.216

6.  Heifetz clip failure. A metallurgical study.

Authors:  M Dujovny; N Kossovsky; R Kossowsky; S Haines; J C Maroon
Journal:  J Neurosurg       Date:  1979-03       Impact factor: 5.115

7.  The broken clip.

Authors:  G F Edner; K Ericson; D M Forster; D M Steiner
Journal:  Acta Neurochir (Wien)       Date:  1978       Impact factor: 2.216

  7 in total
  1 in total

1.  Per-operative angiography--a useful tool in cerebral aneurysm surgery?

Authors:  J Hillman; I Johansson
Journal:  Acta Neurochir (Wien)       Date:  1987       Impact factor: 2.216

  1 in total