| Literature DB >> 6736161 |
Abstract
Porphyrins were separated using ion-pair reversed-phase high-performance liquid chromatography. The eluents were aqueous potassium phosphate buffer and tetrabutylammonium phosphate in methanol. The influences of pH value and ionic strength of the phosphate buffer and molarity of the ion-pair reagent in methanol were investigated to improve separation and detection. A linear response curve was obtained from 0.38 to 7.64 pmol for coproporphyrin I. The detection limits were determined to be 0.12 pmol for coproporphyrin I and 0.22 pmol for uroporphyrin I.Entities:
Mesh:
Substances:
Year: 1984 PMID: 6736161 DOI: 10.1016/s0021-9673(01)93575-7
Source DB: PubMed Journal: J Chromatogr