Evaluation of two devices for improving thermal control of premature infants in transport.

M H LeBlanc.   

Abstract

Two devices, a plastic heat shield and a chemically heated mattress, were evaluated for their ability to reduce the needed air temperature and the power requirements in air servo-controlled transport incubators. A single-walled transporter and a double-walled transporter were evaluated using a heated manikin modeled after a 1000-g baby. The plastic heat shield reduced required transport incubator air temperature by 0.7 degree in the single-walled and 0.5 degree in the double-walled incubator, but did not significantly reduce power consumption. The chemically heated mattress was more effective in reducing power consumption and required air temperature, but it was difficult to stabilize infant temperatures in an incubator without skin servo-control. The transport incubators tested have wide swings in air temperature even during stable operation, rendering them suboptimal to manage small premature infants.

Entities:  

Mesh:

Year:  1984        PMID: 6734227     DOI: 10.1097/00003246-198407000-00010

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  2 in total

1.  Erenumab for the preventive treatment of chronic migraine complicated with medication overuse headache: an observational, retrospective, 12-month real-life study.

Authors:  Maria Michela Cainazzo; Carlo Baraldi; Anna Ferrari; Flavia Lo Castro; Luca Pani; Simona Guerzoni
Journal:  Neurol Sci       Date:  2021-02-05       Impact factor: 3.307

Review 2.  Double wall versus single wall incubator for reducing heat loss in very low birth weight infants in incubators.

Authors:  N Laroia; D L Phelps; J Roy
Journal:  Cochrane Database Syst Rev       Date:  2007-04-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.