A method to measure the MTF of digital x-ray systems.

R A Sones, G T Barnes.   

Abstract

A method has been devised to accurately measure the modulation transfer function (MTF) of digital x-ray systems up to and, for undersampled systems, beyond the pixel Nyquist frequency (fN). A phantom consisting of an array of parallel tungsten or similar wires is imaged, and discrete Fourier transforms of rows of pixel values are computed. Under suitable conditions of phantom orientation, wire diameter, wire spacing, and image magnification, the envelope of the modulus of the mean Fourier transform represents the system MTF. Experimental results extending beyond fN are presented for an undersampled prototype digital chest x-ray system and shown to be in reasonable agreement with predicted values. Employment of the method with other digital imaging modalities [i.e., computerized tomography (CT) scanners and nuclear magnetic resonance (NMR) units] is also discussed as well as error considerations and practical problems in implementing the method.

Mesh:

Year:  1984        PMID: 6727791     DOI: 10.1118/1.595493

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Accurate MTF measurement in digital radiography using noise response.

Authors:  Andrew Kuhls-Gilcrist; Amit Jain; Daniel R Bednarek; Kenneth R Hoffmann; Stephen Rudin
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

2.  Nonparametric two-dimensional point spread function estimation for biomedical imaging.

Authors:  T D Doukoglou; I W Hunter; R E Kearney
Journal:  Med Biol Eng Comput       Date:  1993-05       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.