A laboratory trial of two new blood-glucose reflectance meters featuring automatic external calibration.

S Walford, P D Home, K G Alberti.   

Abstract

Two new battery-operated digital reflectance meters for blood glucose measurement have been tested for precision and accuracy in a laboratory trial. Reflocheck (Boehringer Company Ltd) is calibrated from a bar code printed on each test strip, Reflolux (Boehringer Company Ltd) from a bar code on acetate film provided with each bottle of BM-Test-Glycemie 20-800R strips, which can also be read visually. Three aqueous standards (3.3, 8.3 and 11.7 mmol/l) were assayed 10 times on each of 10 days. Overall coefficients of variation for Reflocheck were 2.8, 1.7 and 2.0% and for Reflolux 3.1, 1.5 and 2.8% respectively. One hundred and four blood samples measured on the machines gave very high correlations with a manual hexokinase assay: for Reflocheck, y = 1.1x - 0.7, r = 0.99 and for Reflolux, y = 1.07x + 0.1, r = 0.99. In the range 2.2-10 mmol, errors of 1 mmol or more occurred on seven occasions with Reflocheck and eight with Reflolux (n = 80) but none of these errors was of therapeutic significance. The performance of both machines was as good or better than previously published data for similar instruments. The ease of calibration and accuracy of both machines are advantageous as is the use of BM-Test-Glycemie 20-800R by Reflolux which enables a check to be made on visual readings.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6712141     DOI: 10.1177/000456328402100207

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  3 in total

1.  Self-monitoring of the blood glucose level: potential sources of inaccuracy.

Authors:  B Rasaiah
Journal:  Can Med Assoc J       Date:  1985-06-15       Impact factor: 8.262

2.  A Plasmonic Fiber Based Glucometer and Its Temperature Dependence.

Authors:  Jisoo Kim; Changjin Son; Sunjak Choi; Won Jung Yoon; Heongkyu Ju
Journal:  Micromachines (Basel)       Date:  2018-10-05       Impact factor: 2.891

3.  Accuracy of blood glucose concentrations determined by four visually read reagent strips.

Authors:  M P MacManus; R D Neely; P M Bell
Journal:  Ulster Med J       Date:  1988-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.