Intracranial clips: an examination of the devices used for aneurysm surgery.

M Dujovny, N Kossovsky, R Kossowsky, A Perlin, R Segal, F G Diaz, J I Ausman.   

Abstract

A properly functioning aneurysm clip is of paramount importance in the surgical ablation of intracranial aneurysms. An appropriate match between the closing force needed to ablate an aneurysm and the force exerted by the aneurysm clip must be achieved. In this study, the opening and closing forces exerted by several commonly used aneurysm clips were tested. There was a strong correlation between the classification of the clips, according to the Dujovny system, and their mechanical behavior. Minimal variability among different lots of the same clip type and minimal hysteresis were generally exhibited by the alpha mobile fulcrum class clips, such as the Yasargil, Vari-Angle McFadden, Scoville, and Mayfield clips. Significant exceptions included several types of Mayfield and Drake clips. In contrast, pivot class clips generally showed twice as much lot variability among different lots, as well as a significant amount of hysteresis. Clips of this class include the Vari-Angle, Heifetz, and Pivot clips. Because significant variations in clip force exist and because several other factors can compound these differences, it is suggested that aneurysm clips be individually tested for their closing forces before they are used in an operation.

Entities:  

Mesh:

Year:  1984        PMID: 6709149     DOI: 10.1227/00006123-198403000-00001

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  4 in total

1.  Revisit of aneurysm clip closing forces: comparison of titanium versus cobalt alloy clip.

Authors:  Tetsuyoshi Horiuchi; Nunung Nur Rahmah; Takao Yanagawa; Kazuhiro Hongo
Journal:  Neurosurg Rev       Date:  2012-06-15       Impact factor: 3.042

2.  Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems.

Authors:  Frank G Shellock; Jean A Tkach; Paul M Ruggieri; Thomas J Masaryk; Peter A Rasmussen
Journal:  AJNR Am J Neuroradiol       Date:  2003-03       Impact factor: 3.825

3.  Scissoring of a Cobalt Alloy Aneurysm Clip causing Slippage during Cerebral Aneurysm Surgery: Case report and review of literature.

Authors:  Rajeev Kariyattil; Dilip Panikar
Journal:  Sultan Qaboos Univ Med J       Date:  2013-02-27

4.  Use of a microsurgical vascular clip system for temporary bilateral occlusion of the four main uterine vessels for laparoscopic enucleation of very large intramural uterine fibroids.

Authors:  Shadi Younes; Marc Radosa; Achim Schneider; Julia Radosa; Alexey Eichenwald; Christiane Weisgerber; Bahriye Aktas
Journal:  Arch Gynecol Obstet       Date:  2022-07-27       Impact factor: 2.493