Histological evaluation of porous titanium cardiac pacemaker electrode tips.

M S Hirshorn, L K Holley, D K Money, M Spector, F A Young, J R Hales.   

Abstract

An investigation of the tissue reaction to five porous titanium alloy pacemaker electrode tips inserted in sheep hearts was made. Attachment of the electrode tips to the endocardial wall was facilitated by fibrous tissue ingrowth into the porous material. Vascularized fibrocollagenous tissue was found within the pores of the tips. One tip was found to be infected. The presence of occasional multinucleated foreign-body giant cells within the porosity was related to the particulate features of the porous tip. The electrode tips had a reduced sensing impedance compared with conventional solid platinum tips. The present findings are consistent with previous observations relating to porous cobalt-chromium alloy and porous platinum-iridium alloy tips and suggest that porous titanium alloy may be a suitable pacemaker electrode tip material.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6699032     DOI: 10.1002/jbm.820180107

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  2 in total

1.  On the formation of fibrous capsule and fluid space around machined and porous blood plasma clot coated titanium.

Authors:  E Jansson; M Källtorp; A Johansson; P Tengvall; P Thomsen
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

2.  Changes in biphasic electrode impedance with protein adsorption and cell growth.

Authors:  Carrie Newbold; Rachael Richardson; Rodney Millard; Christie Huang; Dusan Milojevic; Robert Shepherd; Robert Cowan
Journal:  J Neural Eng       Date:  2010-09-14       Impact factor: 5.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.