A M Buko, L R Phillips, B A Fraser.
Abstract
Conditions were established for analyzing as little as 5 pmol of an underivatized peptide delivered in a glycerol sample matrix as a thin film onto a gold-plated copper sample stage and then bombarded with xenon fast atoms. Calibration of the fast atom bombardment high field mass spectrometer and data system was achieved using cesium iodide/glycerol as a reference. Calibration at several accelerating potentials permitted a mass range from 393 to 5941 u. Several factors were examined that contribute to the quality of the mass spectrum: components within the glycerol such as other peptides, alkali salts, acid and reducing agents; the nature of the fast atom gas; concentration of the peptide delivered to the sample stage; and the effect of the sample stage and sample matrix on sensitivity.Entities:
Mesh:
Substances:
Year: 1983 PMID: 6683112 DOI: 10.1002/bms.1200100504