| Literature DB >> 6636261 |
K S Schwenzer, C H Wang, J P Anhalt.
Abstract
We have extended fluorescence polarization immunoassay (FPIA) technology for the measurement of drugs to include the complex amphoteric glycopeptide antibiotic vancomycin (molecular weight, 1,449). Fluorescein-labeled vancomycin was employed as a tracer, and antisera specific for vancomycin were raised in rabbits by conventional procedures. Tracer, sample, and diluted antiserum were combined, and the polarization of tracer fluorescence is determined in a specially designed fluorometer (Abbott TDx). Because of instrument design, the possibility of fluorescent interferences is minimized. The assay can measure as little as 0.6 mg/L of vancomycin and is free of interferences from hemolysis, lipemia, bilirubin, and changes in protein concentration. The coefficient of variation within assay was 3% (n = 5) and between assays was 5% (n = 5). The FPIA assay (TDx Vancomycin) was compared to a liquid chromatographic (LC) assay for vancomycin and to a commercially available radioimmunoassay (RIA) for 98 clinical specimens. A linear least-squares regression analysis gave a correlation coefficient for LC of 0.980 from the equation FPIA = 1.09 LC + 3.04, and a correlation coefficient for RIA of 0.957 from the equation FPIA = 1.036 RIA + 1.66.Entities:
Mesh:
Substances:
Year: 1983 PMID: 6636261 DOI: 10.1097/00007691-198309000-00017
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681