B-mode ultrasonography of prosthetic vascular grafts.

G A Gooding, K A Herzog, M W Hedgcock, R L Eisenberg.   

Abstract

The role of contact B-mode ultrasonography in demonstrating prosthetic vascular grafts was assessed in 24 patients (27 grafts). All grafts were well delineated. End-to-side bypass grafts could be separated from end-to-end replacement grafts by their characteristic appearance. Patent and occluded grafts could not be distinguished either in vivo or in vitro. However, B-mode ultrasonography was ideally suited to the detection of other complications of vascular prosthetic grafts: hematoma, infection, and false aneurysm formation.

Entities:  

Mesh:

Year:  1978        PMID: 663175     DOI: 10.1148/127.3.763

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105