Correcting for the dynamic response of a respiratory mass spectrometer.

J H Bates, G K Prisk, T E Tanner, A E McKinnon.   

Abstract

Mass spectrometers produce distorted measurements of gas concentrations because of the time delays and rise times inherent in their responses. Three techniques for numerically correcting such distortion were applied to the acetylene step responses of a Perkin-Elmer MGA1100 mass spectrometer and to simulated data. The techniques investigated were 1) a simple time-delay correction, 2) an exponential model method that assumes a biexponential form for the peak of the impulse response, and 3) a Fourier transform method of deconvolution known as Wiener filtering. The time-delay correction produced an order of magnitude reduction in measurement error. The exponential model method improved on the time-delay correction, and the Wiener filter gave the most accurate corrections in all cases examined.

Entities:  

Mesh:

Year:  1983        PMID: 6629899     DOI: 10.1152/jappl.1983.55.3.1015

Source DB:  PubMed          Journal:  J Appl Physiol Respir Environ Exerc Physiol        ISSN: 0161-7567


  9 in total

1.  A real-time algorithm to improve the response time of a clinical multigas analyser.

Authors:  L Wong; R Hamilton; E Palayiwa; C Hahn
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502

2.  Deconvolution filters to enhance resolution of dense time-of-flight survey spectra in the time-lag optimization range.

Authors:  Dariya I Malyarenko; William E Cooke; Eugene R Tracy; Michael W Trosset; O John Semmes; Maciek Sasinowski; Dennis M Manos
Journal:  Rapid Commun Mass Spectrom       Date:  2006       Impact factor: 2.419

3.  Lung function is unchanged in the 1 G environment following 6-months exposure to microgravity.

Authors:  G Kim Prisk; Janelle M Fine; Trevor K Cooper; John B West
Journal:  Eur J Appl Physiol       Date:  2008-05-07       Impact factor: 3.078

4.  VO2 kinetics in subjects differing in aerobic capacity: investigation by spectral analysis.

Authors:  D Essfeld; U Hoffmann; J Stegemann
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1987

5.  Ventilation heterogeneity measured by multiple breath inert gas testing is not affected by inspired oxygen concentration in healthy humans.

Authors:  Susan R Hopkins; Ann R Elliott; G Kim Prisk; Chantal Darquenne
Journal:  J Appl Physiol (1985)       Date:  2017-03-09

6.  Response characteristics of a mass spectrometer.

Authors:  S J Tavener; S J Withy; E A Harris
Journal:  Med Biol Eng Comput       Date:  1984-11       Impact factor: 2.602

Review 7.  Algorithms, modelling and VO₂ kinetics.

Authors:  Carlo Capelli; Capelli Carlo; Michela Cautero; Cautero Michela; Silvia Pogliaghi; Pogliaghi Silvia
Journal:  Eur J Appl Physiol       Date:  2010-02-27       Impact factor: 3.078

8.  Protective effect of prone posture against hypergravity-induced arterial hypoxaemia in humans.

Authors:  M Rohdin; J Petersson; M Mure; R W Glenny; S G E Lindahl; D Linnarsson
Journal:  J Physiol       Date:  2003-02-21       Impact factor: 5.182

9.  Nitrogen-based lung clearance index: a valid physiological biomarker for the clinic.

Authors:  Chantal Darquenne; Rebecca J Theilmann; Janelle M Fine; Sylvia A B Verbanck
Journal:  J Appl Physiol (1985)       Date:  2022-04-21
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.