| Literature DB >> 6628847 |
Abstract
Results comparing hearing aid performance utilizing probe microphones with performance measured in a commercial hearing aid test box using an HA-2 coupler are presented. Sixteen hearing-impaired subjects (23 ears) were divided into two groups. Group I utilized unvented earmolds with short canals and large bores; group II utilized unvented earmolds with long canals and small bores. Results revealed that large individual differences across frequencies as well as within and between groups preclude predicting how a hearing aid will perform on a real ear from test box measurements.Entities:
Mesh:
Year: 1983 PMID: 6628847 DOI: 10.1097/00003446-198309000-00003
Source DB: PubMed Journal: Ear Hear ISSN: 0196-0202 Impact factor: 3.570