Literature DB >> 6619947

A cannula for use in ultrasonically guided biopsies of the brain. Technical note.

J M Rubin, G J Dohrmann.   

Abstract

The authors describe a cannula that has been modified to improve its visibility during ultrasonically guided biopsies of the brain. To enhance the echogenicity of the tip of the cannula, six parallel rings, 0.25 to 0.30 mm deep and 0.40 to 0.50 mm apart, were etched into the tip of the cannula. The cannula was also lengthened to approximately 19 cm to permit its unencumbered passage through any biopsy guidance device that may be employed. Detailing the precise location of the tip of the biopsy cannula is most important. The tip of this modified probe is much more echogenic and hence more easily seen ultrasonically than it would be otherwise, both in vivo and in vitro. This modified cannula is useful in any ultrasonically guided intracranial biopsy procedure.

Mesh:

Year:  1983        PMID: 6619947     DOI: 10.3171/jns.1983.59.5.0905

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  1 in total

1.  The benefits of navigated intraoperative ultrasonography during resection of fourth ventricular tumors in children.

Authors:  Mohamed A El Beltagy; Mostafa M E Atteya
Journal:  Childs Nerv Syst       Date:  2013-04-23       Impact factor: 1.475

  1 in total