| Literature DB >> 6603467 |
Abstract
Cerebrospinal fluid specimens from patients with suspected meningitis were screened with the Phadebact Haemophilus Test (Pharmacia Diagnostics), with Bactogen (Wampole Laboratories), and by counterimmunoelectrophoresis. With culture-positive fluids, Phadebact coagglutination detected 95%, Bactogen latex agglutination detected 91%, and counterimmunoelectrophoresis detected only 79%. Both agglutination techniques were 25-fold more sensitive than counterimmunoelectrophoresis when tested with dilutions of positive fluids. To obtain specific reactions with the Phadebact reagents it was necessary to heat treat (95 degrees C, 5 min) the fluid; with Bactogen and counterimmunoelectrophoresis this was not necessary.Entities:
Mesh:
Substances:
Year: 1983 PMID: 6603467 PMCID: PMC272791 DOI: 10.1128/jcm.17.6.1005-1008.1983
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948