[Calibration of pressure-measuring foil in biomechanics].

V Hendrich, K H Widmer, E Ficker.   

Abstract

Concerning the calibration of the Prescale foil - a pressure sensitive foil, which shows a monochromatic, pressure depending coloration when loaded - some interesting new facts were ascertained. The coloration of the foil does not depend on the pressure load only, but also on the material of the pressing bodies. Calibration using hard material (high Elastic-Module, e.g. steel) results in a lower maximal optical density compared to soft material (low Elastic-Module, e.g. a hydraulic system) at the same pressure level. Due to wellknown facts of the pressure conditioned in a joint, a hydraulic system is more equivalent to joint conditions. Therefore, a calibration method based on a hydraulic system should be preferred if the Prescale foil is used in biomechanics. Furthermore, the density of coloration depends on the speed of pressure increase, too. A faster load of a particular pressure results in a denser coloration. Pressure increase during calibration shall be the same as during investigation. Both effects are statistically significant.

Mesh:

Year:  1984        PMID: 6523628     DOI: 10.1007/bf02586010

Source DB:  PubMed          Journal:  Unfallchirurgie        ISSN: 0340-2649


  3 in total

1.  [Experimental studies on load-bearing pressure and contact areas in the ankle joint].

Authors:  O Paar; B Rieck; P Bernett
Journal:  Unfallheilkunde       Date:  1983-12

2.  [A new method for the determination of the weight-dependent course of pressure and contact in the articular surface].

Authors:  H J Hehne; E Ficker; W Jantz; D Mahr; H J Schöpf
Journal:  Morphol Med       Date:  1981-10

3.  Load-bearing pattern of the ankle joint. Contact area and pressure distribution.

Authors:  M Kimizuka; H Kurosawa; T Fukubayashi
Journal:  Arch Orthop Trauma Surg       Date:  1980
  3 in total
  1 in total

1.  Physiopathology of the knee joint after distal iliotibial band transfer.

Authors:  H Boszotta; W Helperstorfer; A Jusner; K Hoffmann
Journal:  Arch Orthop Trauma Surg       Date:  1992       Impact factor: 3.067

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.