| Literature DB >> 6518941 |
W H Schuette, S E Shackney, F A Plowman, H W Tipton, C A Smith, M A MacCollum.
Abstract
A multibeam optical detection system has been developed with a high optical efficiency, achieved through a reduction in the number of optical interfaces employed in the system. This reduction is made possible by a combination of employing simple lenses, gluing the objective lens directly upon the face of the flow cuvette and the extraction of only one fluorescence signal from each laser beam. A modified flow chamber is also described that includes fluidic resistance elements for the elimination of most of the electric shielding normally associated with electronic cell volume measurements.Mesh:
Year: 1984 PMID: 6518941 DOI: 10.1002/cyto.990050616
Source DB: PubMed Journal: Cytometry ISSN: 0196-4763