Literature DB >> 6507961

A new test circulatory system for research in cardiovascular engineering.

M Arabia, T Akutsu.   

Abstract

A new test circulatory system (TCS) has been developed for the in vitro testing of artificial hearts (AH) and for research in cardiovascular engineering, when connected to an AH that mimics the natural heart. The TCS is controlled by five variables whereby the slopes of the systemic and pulmonary venous return curves and the mean circulatory pressure can be fixed. It allows us to observe the mutual influence between TCS and AH characteristics and particularly the blood volume distribution, the pressure distribution, and the flowrates in steady-state conditions and (in the near future) also in dynamic conditions. A steady-state mathematical analysis describing the TCS is reported. Numerical results for the human circulatory system at different levels of activity and in physiologic conditions are shown. A first prototype of the TCS has been working for more than one year. The experimental results are in agreement with the mathematical analysis.

Entities:  

Mesh:

Year:  1984        PMID: 6507961     DOI: 10.1007/bf02410289

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  9 in total

1.  The relationship of cardiac output and venous pressure in long surviving calves with total artificial heart.

Authors:  E Hennig; C Grosse-Siestrup; W Krautzberger; H Kless; E S Bücherl
Journal:  Trans Am Soc Artif Intern Organs       Date:  1978

2.  Mock circulation to test pumps designed for permanent replacement of damaged hearts.

Authors:  W J KOLFF
Journal:  Cleve Clin Q       Date:  1959-10

3.  Design of a hydraulic analog of the circulatory system for evaluating artificial hearts.

Authors:  F M Donovan
Journal:  Biomater Med Devices Artif Organs       Date:  1975

Review 4.  Circulation: overall regulation.

Authors:  A C Guyton; T G Coleman; H J Granger
Journal:  Annu Rev Physiol       Date:  1972       Impact factor: 19.318

5.  Mock circulation systems for artificial hearts.

Authors:  G L Mrava
Journal:  Adv Biomed Eng Med Phys       Date:  1970

6.  An artificial arterial system for pumping hearts.

Authors:  N Westerhof; G Elzinga; P Sipkema
Journal:  J Appl Physiol       Date:  1971-11       Impact factor: 3.531

Review 7.  The relationship of cardiac output and arterial pressure control.

Authors:  A C Guyton
Journal:  Circulation       Date:  1981-12       Impact factor: 29.690

8.  The cardiac and vascular factors that determine systemic blood flow.

Authors:  M N Levy
Journal:  Circ Res       Date:  1979-06       Impact factor: 17.367

9.  A new automatically controlled electric TAH.

Authors:  M Arabia; C Franconi; M Guerrisi; A Magrini; N Yamamoto; A Vakamudi; M Drummond; T Akutsu
Journal:  Trans Am Soc Artif Intern Organs       Date:  1980
  9 in total