| Literature DB >> 649732 |
Abstract
The optimisation of conditions for the application of reversed-phase high-performance liquid chromatography (HPLC) to the analysis of steroid products is discussed. The influence of temperature and flow-rate on column efficiency are described. At 60 degrees, the reduced velocity required to give maximum efficiency, is approximately twice that found at room temperature, allowing faster through-put. (hv) curves are reported for 5-micron Spherisorb-ODS and Hypersil-ODS material for corticosteroids at 60 degrees. On-column injection techniques are discussed and data presented, showing that for corticosteroids on-column loop injection can be efficient as septum injection using the techniques described.Entities:
Mesh:
Substances:
Year: 1978 PMID: 649732 DOI: 10.1016/s0021-9673(00)80989-9
Source DB: PubMed Journal: J Chromatogr