Different performance of two commercial nebulizers.

W Arossa, F Quagliotti, M Sala, S Spinaci, G De Candussio.   

Abstract

Two different nebulizers, the DeVilbiss 646 and the Mefar were tested for differences in output characteristics (air flow rate, amount of nebulized liquid and particle size) and it was found that there is a high variability between nebulizers of both brands. In the light of these findings, neither of these devices can be chosen as a standard nebulizer for bronchial provocation.

Mesh:

Substances:

Year:  1984        PMID: 6494601     DOI: 10.1159/000194680

Source DB:  PubMed          Journal:  Respiration        ISSN: 0025-7931            Impact factor:   3.580