Literature DB >> 6492218

A study of external skeletal fixation systems for unstable pelvic fractures.

L E Dahners, R R Jacobs, G Jayaraman, A J Cepulo.   

Abstract

This study was undertaken to determine the feasibility of constructing an anterior pelvic external fixator capable of resisting displacement of vertical shear fractures. Newly designed tapered thread pins for cancellous bone had better bone pin fixation and their greater 6-mm diameter provided more than twice the stiffness of Hoffmann 4-mm pins when tested in clusters. Using these pins and a more rigid anterior frame as one model, various pelvic fixators were tested to determine their resistance to vertical shear forces. These tests determined that this new fixator was 16 times stiffer than a Bonnel single anterior frame and five times stiffer than a double anterior frame (both constructed of Hoffmann components). Extrapolation from the data shows that 13 mm of posterior fracture displacement would occur with loads of one half body weight using the new fixator. Additionally it was noted that Hoffmann frames constructed with two 5-mm pins performed as well as those using three 4-mm pins.

Entities:  

Mesh:

Year:  1984        PMID: 6492218     DOI: 10.1097/00005373-198410000-00002

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  2 in total

1.  When do anterior external or internal fixators provide additional stability in an unstable (Tile C) pelvic fracture? A biomechanical study.

Authors:  E Mcdonald; A A Theologis; P Horst; U Kandemir; M Pekmezci
Journal:  Eur J Trauma Emerg Surg       Date:  2014-12-03       Impact factor: 3.693

2.  Case report: The prone reduction of a sacroiliac disruption with a pelvic C-clamp.

Authors:  Andres Javier Quintero; Ivan S Tarkin; Hans-Christoph Pape
Journal:  Clin Orthop Relat Res       Date:  2008-09-23       Impact factor: 4.176

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.