Literature DB >> 647894

Noninvasive recording of His-Purkinje activity in man by ORS-triggered signal averaging.

W J Wajszczuk, M J Stopczyk, M S Moskowitz, R J Zochowski, T Bauld, P L Dabos, M Rubenfire.   

Abstract

Mobile instrumentation and a clinically applicable method have been developed for external His bundle recording. High gain signal amplification (10)(5) filtering (30--300 HZ) and averaging (128 or 256 consecutive cycles) are used. Acquisition of signals arising in the P-R interval is triggered by the patient's QRS signal at the end of that interval. The precordial bipolar electrodiogram is digitized at 5k HZ with 8 bit resolution and transferred to a 1,024 word, 18 bit signal averager. The averaged signal is then displayed on an oscilloscope and photographed. Good correlations were obtained between direct intracardiac and precordial recordings in experimental animals and in humans. Noise level after averaging was below 0.3 microV, and there was good elimination of asynchronous atrial and ectopic ventricular activity. With averaging of 128 or 256 consecutive cycles, the signal attenuation after propagation to the chest wall was in the range 1:2000 to 1:4000 in comparison with the directly recorded His bundle activity deflections. The noninvasive method may be of value in follow-up of acute and chronic disturbances of atrioventricular conduction, as well as in studies of effects of pharmacologic interventions.

Entities:  

Mesh:

Year:  1978        PMID: 647894     DOI: 10.1161/01.cir.58.1.95

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  7 in total

1.  New algorithm for the detection of the ECG fiducial point in the averaging technique.

Authors:  V Barbaro; P Bartolini; M Fierli
Journal:  Med Biol Eng Comput       Date:  1991-03       Impact factor: 2.602

2.  Noninvasive recording of His bundle activity with an analog delay device in normal subjects and patients with atrioventricular block.

Authors:  Y Ushiyama; H Okajima; Y Sasaki; M Hasegawa
Journal:  Heart Vessels       Date:  1995       Impact factor: 2.037

3.  Enhancing the reliability in the noninvasive measurement of the his bundle magnetic field using a novel signal averaging methodology.

Authors:  Sengottuvel Senthilnathan; Parasakthi Chandrasekaran; Mariyappa Narayanan; Rajesh Patel; Gireesan Katholil; Madhukar P Janawadkar; Radhakrishnan S Thimmakudy; Muralidharan R Thoddi
Journal:  Ann Noninvasive Electrocardiol       Date:  2012-07       Impact factor: 1.468

4.  Noninvasive recording of the His bundle electrogram in neonates.

Authors:  A W Nathan; R S Bexton; A M Levy; N Parmar; A J Camm
Journal:  Pediatr Cardiol       Date:  1984 Jan-Mar       Impact factor: 1.655

5.  Recording of surface His-Purkinje potentials.

Authors:  A Peper; R Jonges; T G Losekoot; C A Grimbergen
Journal:  Med Biol Eng Comput       Date:  1985-07       Impact factor: 2.602

6.  Separation of His-Purkinje potentials from coinciding atrium signals: removal of the P-wave from the electrocardiogram.

Authors:  A Peper; R Jonges; T G Losekoot; C Grimbergen
Journal:  Med Biol Eng Comput       Date:  1982-03       Impact factor: 2.602

Review 7.  Electrocardiologic and related methods of non-invasive detection and risk stratification in myocardial ischemia: state of the art and perspectives.

Authors:  Thomas Huebner; Matthias Goernig; Michael Schuepbach; Ernst Sanz; Roland Pilgram; Andrea Seeck; Andreas Voss
Journal:  Ger Med Sci       Date:  2010-10-11
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.