Literature DB >> 6475580

Corrosion of aneurysm clips: evaluation and clinical implications. Part II: Individual performance.

M Dujovny, R Kossowsky, N Kossovsky, F G Diaz, J I Ausman.   

Abstract

The intracranial fracture of aneurysm clips due to stress corrosion failure is an increasingly common long term postoperative complication. To study the susceptibility of different clips to this process, we evaluated seven aneurysm clips (Mayfield, Yasargil, Vari-Angle McFadden, Scoville, Heifetz, and Pivot) in vitro. Microstructures, hardness, chemical composition, and susceptibility to stress corrosion failure were all measured by standard metallurgical techniques. Certain clips such as Heifetz, Pivot, and Vari-Angle showed a much greater propensity to stress corrosion failure than the others. As a result of our tests, we recommend that only the clips that show high stress corrosion resistance in in vitro testing should be considered for long term in vivo implantation.

Entities:  

Mesh:

Year:  1984        PMID: 6475580     DOI: 10.1007/bf01406875

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  9 in total

1.  [Unusual recurrence of an IC--PC aneurysm after neck clipping--any structural weakness of the clip? (author's transl)].

Authors:  I Hayakawa; A Sasaki; K Fujiwara; T Tsuchida
Journal:  Acta Neurochir (Wien)       Date:  1976       Impact factor: 2.216

2.  Delayed breaking of a Heifetz aneurysm clip. Case report.

Authors:  A Servo; M Puranen
Journal:  J Neurosurg       Date:  1977-09       Impact factor: 5.115

3.  Fractured aneurysm clip. Case report.

Authors:  D O Quest; R W Countee
Journal:  J Neurosurg       Date:  1977-09       Impact factor: 5.115

Review 4.  Compatibility of biomaterials.

Authors:  P G Laing
Journal:  Orthop Clin North Am       Date:  1973-04       Impact factor: 2.472

5.  Tissue reactions to standard neurosurgical metallic implants.

Authors:  J T McFadden
Journal:  J Neurosurg       Date:  1972-05       Impact factor: 5.115

6.  Failure of a Heifetz aneurysm clip.

Authors:  R Kossowsky; M Dujovny; N Kossovsky; Y Keravel
Journal:  J Neurosurg       Date:  1982-08       Impact factor: 5.115

7.  Heifetz clip failure. A metallurgical study.

Authors:  M Dujovny; N Kossovsky; R Kossowsky; S Haines; J C Maroon
Journal:  J Neurosurg       Date:  1979-03       Impact factor: 5.115

8.  The broken clip.

Authors:  G F Edner; K Ericson; D M Forster; D M Steiner
Journal:  Acta Neurochir (Wien)       Date:  1978       Impact factor: 2.216

9.  Interrelationship of aneurysm clips and vascular tissue.

Authors:  T J Rosenbaum; T M Sundt
Journal:  J Neurosurg       Date:  1978-06       Impact factor: 5.115

  9 in total
  1 in total

1.  Per-operative angiography--a useful tool in cerebral aneurysm surgery?

Authors:  J Hillman; I Johansson
Journal:  Acta Neurochir (Wien)       Date:  1987       Impact factor: 2.216

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.