Literature DB >> 6453234

Calibration of some instruments for measuring peak expiratory flow.

A Shaw, J Fisher.   

Abstract

The performance of instruments for measuring peak expiratory flow have been investigated. This evaluation was carried out with two designs of spirometer, two types of pneumotachograph and three versions of the Wright peak flow meter. The appropriate flow profiles were generated by means of a special type of flow calibrator. The errors produced by the instruments varied from zero to about 26%.

Mesh:

Year:  1980        PMID: 6453234     DOI: 10.3109/03091908009161706

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  1 in total

1.  Effects of equipment and technique on peak flow measurements.

Authors:  Thomas Bongers; B Ronan O'Driscoll
Journal:  BMC Pulm Med       Date:  2006-06-20       Impact factor: 3.317

  1 in total