Literature DB >> 6424819

Fixed performance oxygen masks: an evaluation.

S L Hill, P K Barnes, T Hollway, R Tennant.   

Abstract

Fixed performance oxygen masks operate by supplying mixtures of oxygen and air at rates exceeding the inspiratory flow rate of the patient. In this study the oxygen concentration delivered by three fixed performance oxygen masks was determined non-invasively at various inspiratory flow rates. At low inspiratory flow rates all the masks studied acted as fixed performance devices. When the peak inspiratory rate increased the performance of all the masks showed some variability. The change from fixed to variable performance depended on the relation between inspiratory flow rate and the total gas flow delivered by the mask and was independent of the volume of the mask. Hence the use of low volume masks and high oxygen flow rates should produce more consistent results than high volume masks and lower flow rates.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6424819      PMCID: PMC1441120          DOI: 10.1136/bmj.288.6426.1261

Source DB:  PubMed          Journal:  Br Med J (Clin Res Ed)        ISSN: 0267-0623


  7 in total

1.  Variation in performance of oxygen therapy devices. Towards the rational employment of 'The dephlogisticated air described by Priestley'.

Authors:  J M Leigh
Journal:  Ann R Coll Surg Engl       Date:  1973-04       Impact factor: 1.891

2.  Oxygen therapy. Evaluation of various air-entraining masks.

Authors:  S A Friedman; B Weber; W A Briscoe; J P Smith; T K King
Journal:  JAMA       Date:  1974-04-22       Impact factor: 56.272

3.  How to use the Venturi mask.

Authors:  E J Campbell
Journal:  Lancet       Date:  1982-11-27       Impact factor: 79.321

4.  Fixed performance oxygen masks. Hypoxic hazard of low-capacity drugs.

Authors:  D Cox; C Gillbe
Journal:  Anaesthesia       Date:  1981-10       Impact factor: 6.955

5.  An analysis of the performance of a variable venturi-type oxygen mask.

Authors:  D F Woolner; J Larkin
Journal:  Anaesth Intensive Care       Date:  1980-02       Impact factor: 1.669

6.  Measurement of respiratory flow by inductance pneumography.

Authors:  S L Hill; J P Blackburn; T R Williams
Journal:  Med Biol Eng Comput       Date:  1982-07       Impact factor: 2.602

7.  Controlled oxygen therapy at 60% concentration. Why and how.

Authors:  E J Campbell; K B Minty
Journal:  Lancet       Date:  1976-06-05       Impact factor: 79.321

  7 in total
  6 in total

1.  Should we judge a mask by its cover?

Authors:  A A Jeffrey; P M Warren
Journal:  Thorax       Date:  1992-07       Impact factor: 9.139

2.  Accuracy of a novel system for oxygen delivery to small children.

Authors:  Matthew Coghill; Namasivayam Ambalavanan; Robert L Chatburn; Patricia L Hibberd; Linda L Wright; Waldemar A Carlo
Journal:  Pediatrics       Date:  2011-07-04       Impact factor: 7.124

3.  Concentrations of oxygen delivered by air entrainment oxygen masks.

Authors:  J M Goddard
Journal:  Ann R Coll Surg Engl       Date:  1985-11       Impact factor: 1.891

Review 4.  A practical approach to oxygen therapy in cystic fibrosis.

Authors:  M E Dodd; C S Haworth; A K Webb
Journal:  J R Soc Med       Date:  1998       Impact factor: 18.000

5.  The OxyMask(™) development and performance in healthy volunteers.

Authors:  James E Paul; Horia Hangan; Julius Hajgato
Journal:  Med Devices (Auckl)       Date:  2008-12-11

6.  Exome sequencing identifies novel mutation signatures of UV radiation and trichostatin A in primary human keratinocytes.

Authors:  Yao Shen; Wootae Ha; Wangyong Zeng; Dawn Queen; Liang Liu
Journal:  Sci Rep       Date:  2020-03-18       Impact factor: 4.379

  6 in total