Literature DB >> 6421986

Blood ammonia measurement using a simple reflectometer.

R N Ratnaike, J E Buttery, S Hoffmann.   

Abstract

We have assessed a compact Blood Ammonia Checker System (Ammonia Checker) consisting of a reflectometer which measures the intensity of colour formed by blood ammonia on a bromocresol green indicator reagent plate. The results show good correlation (r = 0.95) with our routine chemical method and a regression line y = 0.835X - 1.468. The Ammonia Checker has good precision (CV less than 10%) at the critical blood ammonia concentration and accurately measures predetermined ammonia concentrations. Blood sampling is improved with a precision pipette. The Ammonia Checker is simple, convenient and reliable and is ideally suited for a laboratory required to perform an urgent blood ammonia measurement.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6421986     DOI: 10.1515/cclm.1984.22.1.105

Source DB:  PubMed          Journal:  J Clin Chem Clin Biochem        ISSN: 0340-076X


  5 in total

Review 1.  New uses for old urine tests.

Authors:  H Savolainen
Journal:  Br J Ind Med       Date:  1989-06

2.  Urinary alkoxyacetic acids and renal effects of exposure to ethylene glycol ethers.

Authors:  J Laitinen; J Liesivuori; H Savolainen
Journal:  Occup Environ Med       Date:  1996-09       Impact factor: 4.402

3.  Inhibitory effect of intravenous lysine infusion on urea cycle metabolism.

Authors:  T Kato; M Sano; N Mizutani
Journal:  Eur J Pediatr       Date:  1987-01       Impact factor: 3.183

4.  Effect of renal formic acid excretion on urinary calcium and ammonia concentrations.

Authors:  J Liesivuori; H Savolainen
Journal:  Klin Wochenschr       Date:  1987-09-15

5.  Kinetics and renal effects of formic acid in occupationally exposed farmers.

Authors:  J Liesivuori; J Laitinen; H Savolainen
Journal:  Arch Toxicol       Date:  1992       Impact factor: 5.153

  5 in total