Sources of electron contamination for the Clinac-35 25-MV photon beam.

P L Petti, M S Goodman, J M Sisterson, P J Biggs, T A Gabriel, R Mohan.   

Abstract

A detailed Monte Carlo approach has been employed to investigate the sources of electron contamination for the 25-MV photon beam generated by Varian's Clinac-35. Three sources of contamination were examined: (a) the flattening filter and beam monitor chamber, (b) the fixed primary collimators downstream from the monitor chamber and the adjustable photon jaws, and (c) the air volume separating the treatment head from the observation point. Five source-to-surface distances (SSDs) were considered for a single field size, 28 cm in diameter at 80 cm SSD. It was found that for small SSDs (80-100 cm), the dominant sources of electron contamination were the flattening filter and the beam monitor chamber which accounted for 70% of the unwanted electrons. Thirteen percent of the remaining electrons originated in the downstream primary collimators and the photon jaws, and 17% were produced in air. At larger SSDs, the fraction of unwanted electrons originating in air increased. At 400 cm SSD, 61% of the contaminating electrons present in the beam were produced in air, 34% originated in the flattening filter and beam monitor chamber, and 5% were due to interactions in the fixed collimators downstream from the monitor chamber and the adjustable photon jaws. These calculated results are substantiated by recent experiments.

Mesh:

Year:  1983        PMID: 6419032     DOI: 10.1118/1.595348

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Dose discrepancies in the buildup region and their impact on dose calculations for IMRT fields.

Authors:  Shu-Hui Hsu; Jean M Moran; Yu Chen; Ravi Kulasekere; Peter L Roberson
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  Monte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam.

Authors:  Mohammad Taghi Bahreyni Toossi; Mehdi Momennezhad; Seyed Mohammad Hashemi
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

3.  Effect of various physical parameters on surface and build-up dose for 15-MV X-rays.

Authors:  Girigesh Yadav; R S Yadav; Alok Kumar
Journal:  J Med Phys       Date:  2010-10

Review 4.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

5.  Surface buildup dose dependence on photon field delivery technique for IMRT.

Authors:  Shigeru Yokoyama; Peter L Roberson; Dale W Litzenberg; Jean M Moran; Benedick A Fraass
Journal:  J Appl Clin Med Phys       Date:  2004-04-01       Impact factor: 2.102

6.  Skin dose estimation for various beam modifiers and source-to-surface distances for 6MV photons.

Authors:  Girigesh Yadav; R S Yadav; Alok Kumar
Journal:  J Med Phys       Date:  2009-04

7.  Equivalent square formula for determining the surface dose of rectangular field from 6 MV therapeutic photon beam.

Authors:  Lukkana Apipunyasopon; Somyot Srisatit; Nakorn Phaisangittisakul
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

8.  A Monte Carlo evaluation of beam characteristics for total body irradiation at extended treatment distances.

Authors:  Roumiana Chakarova; Marcus Krantz
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

9.  A Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac.

Authors:  Mohammad Taghi Bahreyni Toossi; Marziyeh Behmadi; Mahdi Ghorbani; Hamid Gholamhosseinian
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

10.  Beam flatness modulation for a flattening filter free photon beam utilizing a novel direct leaf trajectory optimization model.

Authors:  Nicholas J Potter; Guanghua Yan; Hongcheng Liu; Haitham Alahmad; Darren L Kahler; Chihray Liu; Jonathan G Li; Bo Lu
Journal:  J Appl Clin Med Phys       Date:  2020-03       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.