Transcutaneous and capillary pCO2 and pO2 measurements in healthy adults.

P D Wimberley, K G Pedersen, J Thode, N Fogh-Andersen, A M Sørensen, O Siggaard-Andersen.   

Abstract

Transcutaneous carbon dioxide and oxygen tensions (tc-pCO2 and tc-pO2) were measured in seven healthy adult volunteers during hyperventilation in atmospheric air and during CO2 inhalation. Three skin sensors were applied to each subject: an O2 electrode, a CO2 electrode, and a combined O2-CO2 electrode, each heated to 44 degrees C. We observed close correlation between tc-pCO2 and capillary-pCO2, the relation being close to that calculated from the anaerobic temperature coefficient of pCO2 in blood. For O2, on the other hand, the relationship between transcutaneous and capillary values appeared more complex. Electrode drift during in-vivo monitoring was greater for pCO2 (up to 12%) than for pO2 (up to 7%), but generally we observed no differences in drift between the combined and the single electrodes. We conclude that tc-pCO2 measured with a single or a combined electrode reliably predicts capillary-pCO2 in healthy adults and that changes are rapidly observed. Our conclusions regarding tc-pO2 values are less definite because of uncertain interpretation of the capillary-pO2 values.

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6409449

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  7 in total

1.  New transcutaneous pO2 probe based on mass spectrometry with low stirring effect.

Authors:  J Grønlund; O Johansen; B Jensen
Journal:  Med Biol Eng Comput       Date:  1985-05       Impact factor: 2.602

2.  Comparison of arterial, end-tidal and transcutaneous PCO2 during moderate exercise and external CO2 loading in humans.

Authors:  U Hoffmann; D Essfeld; J Stegemann
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1990

3.  Effects of the electrode temperature of a new monitor, TCM4, on the measurement of transcutaneous oxygen and carbon dioxide tension.

Authors:  Tomoki Nishiyama; Shinji Nakamura; Koichi Yamashita
Journal:  J Anesth       Date:  2006       Impact factor: 2.078

4.  Noninvasive monitoring of carbon dioxide: a comparison of the partial pressure of transcutaneous and end-tidal carbon dioxide with the partial pressure of arterial carbon dioxide.

Authors:  C Q Phan; K K Tremper; S E Lee; S J Barker
Journal:  J Clin Monit       Date:  1987-07

5.  Measurement of tissue oxygen tension: comparison between two subcutaneous oxygen tonometers.

Authors:  H O Soini; J Takala
Journal:  J Clin Monit       Date:  1991-07

6.  Evaluation of time courses of agreement between minutely obtained transcutaneous blood gas data and the gold standard arterial data from spontaneously breathing Asian adults, and various subgroup analyses.

Authors:  Akira Umeda; Masahiro Ishizaka; Masamichi Tasaki; Tateki Yamane; Taiji Watanabe; Yasushi Inoue; Taichi Mochizuki; Yasumasa Okada; Sarah Kesler
Journal:  BMC Pulm Med       Date:  2020-05-29       Impact factor: 3.317

7.  Bioluminescence, chemiluminescence.

Authors:  A Dwenger; C Beychok; G Schweitzer; H C Pape; G Röllig; M L Nerlich; E Jonas; M Funck; T Zimmermann; S Albrecht; R Schuster; G Lauschke; W Jaroß; V Kaever; E Schmitz; K Resch; H Brandl; W-D Böhm; R Beckert; E Köstler; M Menschikowski; D Kacian; T Lawrence; M Sanders; J Putnam; M Majlessi; S McDonough; T Ryder; J J Santana Rodríguez; Z Sosa Ferrera; A Afonso Perera; V González Díaz
Journal:  Fresenius J Anal Chem       Date:  1990
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.