Evaluation of a rapid bacteriuria screening instrument.

J S Cashman, P A Ahlin, J M Matsen.   

Abstract

A newly modified, semiautomated instrument (Bacteriuria detection device (BDD] designed to detect the presence of bacteriuria in less than 3 min was compared to quantitative urine culture plating techniques. The instrument consists of a self-enclosed vacuum-filtration-staining system in which a 1-ml urine filtrate is stained on a filter. The resulting color determines the quantitation. Of the total of 525 clinical urine specimens tested, 66 (12.6%) were uninterpretable due to pigment deposition or inability to complete the filtering process (clogging of the filter). Of the remaining 459 specimens, 93 (20.3%) had a plate quantitation colony count of 10(5) colony-forming units (CFU)/ml or more. The BDD detected 94.2% of these positive specimens if only significant pathogens were included (85% if specimens that were probably contaminated were also included). For specimens containing significant pathogens at 10(4)-10(5) CFU/ml, the BDD detection rate was 41%. The device detected most (94.3%) gram-negative bacilli and enterococci at colony counts of 10(5) CFU/ml or more. In addition, the BDD accurately detected 95.6% of specimens with no growth or fewer than 10(4) CFU/ml. With several proposed modifications, these results suggest that this instrument is potentially useful as a urine screening device in a select population.

Entities:  

Mesh:

Year:  1984        PMID: 6399478     DOI: 10.1016/0732-8893(84)90018-x

Source DB:  PubMed          Journal:  Diagn Microbiol Infect Dis        ISSN: 0732-8893            Impact factor:   2.803


  2 in total

1.  Evaluation of three bacteriuria screening methods in a clinical research hospital.

Authors:  T C Wu; E C Williams; S Y Koo; J D MacLowry
Journal:  J Clin Microbiol       Date:  1985-05       Impact factor: 5.948

Review 2.  Detection of urinary tract infections by rapid methods.

Authors:  M Pezzlo
Journal:  Clin Microbiol Rev       Date:  1988-07       Impact factor: 26.132

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.