| Literature DB >> 6394622 |
M K Sylvester, J A Washington.
Abstract
The Quantum II Microbiology System correctly identified 250 (89.6%) of 279 gram-negative bacilli from clinical specimens. There were 13 (4.7%) errors in genus identification and 7 (2.5%) errors in species identification; another nine (3.2%) isolates yielded biochemical reaction patterns which were not in the file of the system. Total test time per isolate, excluding inoculum preparation and incubation, was ca. 1.5 min.Mesh:
Year: 1984 PMID: 6394622 PMCID: PMC271548 DOI: 10.1128/jcm.20.6.1196-1197.1984
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948