Literature DB >> 6394622

Evaluation of the Quantum II Microbiology System for bacterial identification.

M K Sylvester, J A Washington.   

Abstract

The Quantum II Microbiology System correctly identified 250 (89.6%) of 279 gram-negative bacilli from clinical specimens. There were 13 (4.7%) errors in genus identification and 7 (2.5%) errors in species identification; another nine (3.2%) isolates yielded biochemical reaction patterns which were not in the file of the system. Total test time per isolate, excluding inoculum preparation and incubation, was ca. 1.5 min.

Mesh:

Year:  1984        PMID: 6394622      PMCID: PMC271548          DOI: 10.1128/jcm.20.6.1196-1197.1984

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  1 in total

1.  Evaluation of the Repliscan system for identification of Enterobacteriaceae.

Authors:  S D Brown; J A Washington
Journal:  J Clin Microbiol       Date:  1978-12       Impact factor: 5.948

  1 in total
  6 in total

Review 1.  Automated systems for identification of microorganisms.

Authors:  C E Stager; J R Davis
Journal:  Clin Microbiol Rev       Date:  1992-07       Impact factor: 26.132

2.  Evaluation of Spectrum-10 system for identification of members of the family Enterobacteriaceae.

Authors:  A Vuye
Journal:  J Clin Microbiol       Date:  1989-10       Impact factor: 5.948

3.  Packaged identification systems and bacteria of veterinary origin.

Authors:  P J Blackwell
Journal:  J Clin Microbiol       Date:  1988-04       Impact factor: 5.948

4.  Evaluation of Quantum II microbiology system for identification of gram-negative bacteria of veterinary origin.

Authors:  R L Jones; W S Adney; M A Davis; H Vonbyren; G Thompson
Journal:  J Clin Microbiol       Date:  1987-11       Impact factor: 5.948

5.  Comparison of the Quantum II Bacterial Identification System and the AutoMicrobic System for the identification of gram-negative bacilli.

Authors:  M A Pfaller; M J Bale; K R Schulte; F P Koontz
Journal:  J Clin Microbiol       Date:  1986-01       Impact factor: 5.948

6.  Critical evaluation of the AutoMicrobic system gram-negative identification card for identification of glucose-nonfermenting gram-negative rods.

Authors:  J J Plorde; J A Gates; L G Carlson; F C Tenover
Journal:  J Clin Microbiol       Date:  1986-02       Impact factor: 5.948

  6 in total