J L Nxumalo, M Teranaka, W G Schenk.
Abstract
The validity of the ICG clearance method for the measurement of THBF in abnormal circulatory states remains questionable. In this study THBF measured by this method is compared with the electromagnetic flow technique in a canine spetic model. Good correlation is demonstrated between the two in normal control animals. However, in the septic animals the ICG underestimated the electromagnetic flow result by 20%. This is true in both the high and the low cardiac output septic shock pictures that emerge. In the septic animals, the total hepatic blood flow as measured by the ICG was almost equal to the portal vein flow alone measured by the electromagnetic flowmeters suggesting that this was the quantity it was measuring in this abnormal state. Pathophysiologic mechanisms that may explain the discrepancy are given.Entities:
Mesh:
Substances:
Year: 1978 PMID: 637587 PMCID: PMC1396436 DOI: 10.1097/00000658-197803000-00015