J F Polak, B L Holman, J L Moretti, R L Eisner, J Lister-James, R J English.
Abstract
The performance of a slant-hole collimator was compared with that of a standard straight-bore, low-energy collimator for tomographic imaging of I-123-iodinated amine brain agents. Improved in-slice resolution was due to the greater proximity between collimator and the subjects' heads. We conclude that high quality tomographic images of the brain can be obtained from rotating cameras equipped with slant-hole collimators.Entities:
Mesh:
Substances:
Year: 1984 PMID: 6336218
Source DB: PubMed Journal: J Nucl Med ISSN: 0161-5505 Impact factor: 10.057