Literature DB >> 6308107

[Studies on enzymeimmunoassay for human chorionic gonadotropin and its clinical application].

O Kobayashi, S Shirotake, H Takamizawa.   

Abstract

Two enzyme immunoassays for hCG were developed and compared with HAR and RIA by using many urine and serum samples. The EIA was characterized by management of trophoblastic disease. 1. An EIA by hCG-beta antibody immobilized on a microplate and hCG antibody conjugated with HRP enzyme quantified hCG from 0.5 to 100ng/ml, while the other by biotinyl hCG antibody and avidine-HRP conjugate quantified hCG from 0.1 to 20ng/ml. Both of them had a good correlation with RIA; y = 1.02x + 2.3, r=0.99 and Y=0.96x-0.38. The C.V.S. were 1.8 and 5.2% (intraassay), or 5.8 and 9.5% (interassay), respectively. 2. The EIA was carried out to test hCG patients whose urinary hCG by HAR was under the LH level. The values in 371 samples were under 2ng/ml, but those in 40 samples were higher than 2ng/ml (Max 15.8 ng/ml). Diagnosis by EIA distinguished recurrence from remission 2 month earlier than by HAR. The EIA was useful in the management of the disease.

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6308107

Source DB:  PubMed          Journal:  Nihon Sanka Fujinka Gakkai Zasshi        ISSN: 0300-9165


  1 in total

1.  Enzyme immunoassay for placental protein 4 (PP4) and its possible diagnostic significance in patients with genital tract cancer.

Authors:  Y Ota; N Inaba; S Shirotake; I Fukazawa; H Takamizawa; H Bohn
Journal:  Arch Gynecol Obstet       Date:  1990       Impact factor: 2.344

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.