Magnetocardiograms of patients with left ventricular overloading recorded with a second-derivative SQUID gradiometer.

K Fujino, M Sumi, K Saito, M Murakami, T Higuchi, Y Nakaya, H Mori.   

Abstract

Magnetocardiograms (MCGs) were recorded by means of a second-derivative SQUID (superconducting quantum interference device) magnetometer in 60 normal subjects and 95 patients with left ventricular overloading to determine the clinical value of the MCG. In patients with left ventricular overloading, the Q or S wave was increased in the upper anterior part of the thorax, and the R wave was increased in the left lower part of the thorax, indicating increased leftward force due to left ventricular overloading. For detection of left ventricular hypertrophy or dilatation from echocardiographic measurements, the sensitivity and specificity of the MCG were similar to those of the standard ECG, or slightly better. In patients with left ventricular systolic overloading, the Q wave was decreased in the lower anterior part of the thorax, indicating a decreased septal vector. Inversion of the T wave was seen more frequently in the MCGs than in the ECGs of patients with left ventricular overloading, suggesting that the MCG is useful for detecting early abnormalities of repolarization. These results suggest that the MCG may provide information that is difficult to obtain from the standard 12-lead ECG.

Entities:  

Mesh:

Year:  1984        PMID: 6237164     DOI: 10.1016/s0022-0736(84)80058-8

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  2 in total

1.  Detection of repolarization abnormalities in patients with cardiomyopathy using current vector mapping technique on magnetocardiogram.

Authors:  Junko Shiono; Hitoshi Horigome; Akira Matsui; Yasushi Terada; Tsuyoshi Miyashita; Keiji Tsukada
Journal:  Int J Cardiovasc Imaging       Date:  2003-04       Impact factor: 2.357

2.  Clinical magnetocardiography: experience with a biomagnetic multichannel system.

Authors:  W Moshage; S Achenbach; A Weikl; K Göhl; K Bachmann; K Abraham-Fuchs; W Härer; S Schneider
Journal:  Int J Card Imaging       Date:  1991
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.