Literature DB >> 6209684

Implantation of automatic cardioverter-defibrillators via median sternotomy.

R Brodman, J D Fisher, S Furman, D R Johnston, S G Kim, J A Matos, L E Waspe.   

Abstract

15 AICD (automatic implantable cardioverter-defibrillator) Model B units were implanted in 10 patients. The median sternotomy is our preferred surgical approach using a right atrial patch electrode, a left ventricular apex patch electrode, and two closely placed epicardial sensing electrodes. Follow-up is 109 patient months and all patients are alive. AICD units discharged for ventricular tachycardia, ventricular flutter, and ventricular fibrillation. Discharges also occurred for sinus tachycardia and atrial fibrillation above the rate limit in three units. Premature pulse generator depletion has occurred in four AICD-B units 3 to 18 months postimplant and appears due to a defect in original battery design. Discharge of the AICD for supraventricular tachycardia is a problem that will remain until a better means of differentiating supraventricular tachycardia from ventricular tachyarrhythmias is found. The AICD appears to prevent sudden death from ventricular tachyarrhythmias.

Entities:  

Mesh:

Year:  1984        PMID: 6209684     DOI: 10.1111/j.1540-8159.1984.tb05709.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

Review 1.  Current state of knowledge and experts' perspective on the subcutaneous implantable cardioverter-defibrillator.

Authors:  Massimo Santini; Riccardo Cappato; Dietrich Andresen; Johannes Brachmann; D Wyn Davies; John Cleland; Alessandro Filippi; Edoardo Gronda; Richard Hauer; Gerhard Steinbeck; David Steinhaus
Journal:  J Interv Card Electrophysiol       Date:  2009-03-06       Impact factor: 1.900

Review 2.  Subcutaneous chronic implantable defibrillation systems in humans.

Authors:  Riccardo Cappato; Warren M Smith; Margaret A Hood; Ian G Crozier; Luc Jordaens; Stefan G Spitzer; Andrey V Ardashev; Lucas Boersma; Pierpaolo Lupo; Andrew A Grace; Gust H Bardy
Journal:  J Interv Card Electrophysiol       Date:  2012-03-17       Impact factor: 1.900

  2 in total