Literature DB >> 6209650

Pacemaker syndrome with DDD pacing.

J Torresani, A Ebagosti, G Allard-Latour.   

Abstract

DDD pacing may be associated with certain problems similar to those of the pacemaker syndrome. In the absence of mechanical dysfunction, pacemaker-mediated tachycardias have occurred in the presence of VA conduction. The correction of this problem is found in more versatile programming capability. In addition, certain phenomena relating to intraatrial conduction times and the optimal AV delay are discussed.

Entities:  

Mesh:

Year:  1984        PMID: 6209650     DOI: 10.1111/j.1540-8159.1984.tb05675.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

Review 1.  Pacemaker syndrome: an iatrogenic condition.

Authors:  C M Travill; R Sutton
Journal:  Br Heart J       Date:  1992-08

2.  Recommendations for pacemaker prescription for symptomatic bradycardia. Report of a working party of the British Pacing and Electrophysiology Group.

Authors: 
Journal:  Br Heart J       Date:  1991-08

3.  Pacemaker syndrome.

Authors:  R A Kenny; R Sutton
Journal:  Br Med J (Clin Res Ed)       Date:  1986-10-11

4.  "Subclinical" pacemaker syndrome: a randomised study of symptom free patients with ventricular demand (VVI) pacemakers upgraded to dual chamber devices.

Authors:  N Sulke; A Dritsas; J Bostock; A Wells; R Morris; E Sowton
Journal:  Br Heart J       Date:  1992-01
  4 in total