Spinal cord protection during cross-fire irradiation of the intrathoracic esophagus: tube-tilt vs. Shielding.

W Swechand, T K Jones, F M Khan, S H Levitt.   

Abstract

The protection of the upper thoracic spine achieved by a tube-tilt technique was found to be comparable to that obtained from direct shielding during cross-fire therapy of the intrathoracic esophagus. An isocentric technique was utilized which included one anterior and two posterior-oblique fields. Significant underdosages to the esophagus at levels superior to the isocenter are caused by tube-tilt, whereas direct shielding resulted in a uniformity of esophageal doses to within 5%. The need for a tissue compensator for the anterior field must be assessed on an individual basis from off-axis treatment planning.

Mesh:

Year:  1978        PMID: 619417     DOI: 10.1148/126.1.239

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105