| Literature DB >> 6180387 |
R A Freedman, M T Rothman, J W Mason.
Abstract
Atrial synchronous pacemakers have been known to cause a variety of cardiac arrhythmias. Of particular concern are those arrhythmias involving a pacemaker stimulus occurring on a T wave, because these may lead to ventricular tachycardia. The Medtronic 2409 ASVIP pacemaker is an atrial synchronous pacemaker with several features designed to decrease the likelihood of such arrhythmias. We report a patient in whom a normally functioning Medtronic 2409 ASVIP pacemaker, despite these features, induced recurrent ventricular tachycardia. Conditions which predisposed this patient to pacemaker-induced re-entrant arrhythmias are discussed.Entities:
Mesh:
Substances:
Year: 1982 PMID: 6180387 DOI: 10.1111/j.1540-8159.1982.tb02266.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976