Endless loop tachycardia in an AV universal [DDD] pacemaker.

S Furman, J D Fisher.   

Abstract

Implantation of an AV universal [DDD] pacemaker, stimulating and sensing both atrium and ventricle caused a pacemaker mediated tachycardia, begun by a P wave occurring after a premature ventricular contraction, and sustained by P waves caused by retrograde AV conduction. The tachycardia was terminated by application of a magnet to the pulse generator which temporarily made it insensitive to incoming signals, both P and QRS.

Entities:  

Mesh:

Year:  1982        PMID: 6180386     DOI: 10.1111/j.1540-8159.1982.tb02265.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Clinically relevant computer model of cardiac rhythm and pacemaker/heart interaction.

Authors:  M Malík; T Cochrane; D W Davies; A J Camm
Journal:  Med Biol Eng Comput       Date:  1987-09       Impact factor: 2.602

2.  Bigeminy and a Pacemaker.

Authors:  Luke Cunningham; Yochai Birnbaum
Journal:  Tex Heart Inst J       Date:  2017-08-01

3.  Effects of cardioactive medications on retrograde conduction: continuing relevance for current devices.

Authors:  John E O'Mara; John D Fisher; Sue J Goldie; Soo G Kim; Kevin J Ferrick; Jay N Gross; Eugen C Palma
Journal:  J Interv Card Electrophysiol       Date:  2006-01       Impact factor: 1.759

4.  Pacemaker-mediated tachycardia with narrow QRS complexes.

Authors:  Hyun Sok Yoo; Ernesto Albino; Pablo Fernandez; Daniel Etcheverry; Rita Tepper; Hugo Garro; Julio Pastori; Rafael Acunzo
Journal:  Clin Case Rep       Date:  2018-04-14

5.  Clinical and Electrophysiological Predictors of Ventriculoatrial Conduction in Patients under Evaluation for Ventricular Tachyarrhythmia.

Authors:  Mohammad Reza Dehghani; Azam Soleimani; Mohammad Reza Samiei Nasab
Journal:  J Tehran Heart Cent       Date:  2010-09-30
  5 in total