Implantation of an AV universal [DDD] pacemaker, stimulating and sensing both atrium and ventricle caused a pacemaker mediated tachycardia, begun by a P wave occurring after a premature ventricular contraction, and sustained by P waves caused by retrograde AV conduction. The tachycardia was terminated by application of a magnet to the pulse generator which temporarily made it insensitive to incoming signals, both P and QRS.Entities:
Mesh:
Year: 1982 PMID: 6180386 DOI: 10.1111/j.1540-8159.1982.tb02265.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976