Literature DB >> 6176965

Pacer-induced tachycardia associated with an atrial synchronous ventricular inhibited (ASVIP) pulse generator.

A O Tolentino, R P Javier, C Byrd, P Samet.   

Abstract

A 68-year-old white male underwent permanent pacemaker implantation with an atrial synchronous ventricular inhibited pulse generator (Medtronic model 2409) because of syncope and abnormal H-V interval of 70 ms. Paroxysmal bouts of pacemaker associated tachycardia were subsequently recorded on several occasions, initiated and terminated by spontaneous ventricular premature beats. The mechanism for the occurrence of the tachyarrhythmia is discussed in detail and the functional characteristics of the pulse generator are described. Replacement of the unit with a different pacer device prevented further occurrence of the arrhythmia.

Entities:  

Mesh:

Year:  1982        PMID: 6176965     DOI: 10.1111/j.1540-8159.1982.tb02221.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Bigeminy and a Pacemaker.

Authors:  Luke Cunningham; Yochai Birnbaum
Journal:  Tex Heart Inst J       Date:  2017-08-01
  1 in total