Short-term threshold behavior of human ventricular pacing electrode: noninvasive monitoring with a multiprogrammable pacing system.

D P Scoblionko, E L Rolett.   

Abstract

Twenty-two patients were studied for 8 to 12 weeks following pacemaker implantation. A unipolar multiprogrammable pacing system was used to assess and compensate for early post-implant changes in pulse width threshold (PWT). Changes in PWT were estimated by noninvasive measurements of the lowest pulse width required for 100% capture at constant voltage. All patients showed an early increase in PWT. The magnitude of this change was unpredictable and ranged from 2- to 10-fold. For pacing, we selected an appropriate margin of safety to be a pulse width three times the PWT. During the study period, 4 patients showed critical increases in PWT, requiring reprogramming to an increased pulse width as above, whereas only one patient required lead repositioning. The majority of patients achieved a stable low PWT and were reprogrammed to a reduced pulse width. The programmable pacing system provides a safe accurate technique for monitoring early changes in PWT, defines the need for appropriate alterations in programmable functions, decreases the risk of early pacemaker failure, and offers promise for improved pacemaker longevity in many patients.

Entities:  

Mesh:

Year:  1981        PMID: 6173852     DOI: 10.1111/j.1540-8159.1981.tb06245.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Bradycardia-dependent rise in the atrial capture threshold early after cardiac pacemaker implantation in patients with sick sinus syndrome.

Authors:  Akira Kimata; Kentaro Yoshida; Noriyuki Takeyasu; Rieko Nakagami; Jun Osada; Takeshi Mitsuhashi; Kazutaka Aonuma; Akihiko Nogami
Journal:  HeartRhythm Case Rep       Date:  2015-10-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.