Abstract
A rapid and reproducible electrophoretic method for the separation and quantitation of haemoglobins on cellulose acetate is described. The accuracy of the method and its possible sources of error are discussed. The normal range for haemoglobin A(2) by this method is 1% to 3% of the total haemoglobin concentration. Blood samples from 32 thalassaemic patients showed haemoglobin A(2) values of 3.5% to 7%.Entities:
Mesh:
Year: 1965 PMID: 5844210 PMCID: PMC473136 DOI: 10.1136/jcp.18.6.790