Safety precautions to be observed with cooled premixed gases.

A Bracken, G B Broughton, D W Hill.   

Abstract

The gas delivered from a cylinder of Entonox (50% nitrous oxide and 50% oxygen) may not contain adequate quantities of oxygen if the cylinder has been cooled and then allowed to warm up in a vertical position.The composition of gas delivered from cylinders cooled to not lower than -40 degrees C. remains unaltered if the cylinders are rewarmed for 24 hours in a horizontal position at a temperature of not less than 5 degrees C.Cylinders of Entonox delivered in cold weather should be stored in a horizontal position at 5 degrees C. or more for at least 24 hours before use.

Entities:  

Mesh:

Substances:

Year:  1968        PMID: 5673962      PMCID: PMC1989594          DOI: 10.1136/bmj.3.5620.715

Source DB:  PubMed          Journal:  Br Med J        ISSN: 0007-1447


  5 in total

1.  EVALUATION OF THE SERVOMEX PARAMAGNETIC ANALYZER.

Authors:  J F NUNN; N A BERGMAN; A J COLEMAN; D C CASSELLE
Journal:  Br J Anaesth       Date:  1964-10       Impact factor: 9.166

2.  PREMIXED GAS AND OXYGEN FOR MIDWIVES.

Authors:  C W GALE; M E TUNSTALL; C C WILTON-DAVIES
Journal:  Br Med J       Date:  1964-03-21

3.  EFFECT OF COOLING ON PRE-MIXED GAS MIXTURES FOR OBSTETRIC ANALGESIA.

Authors:  M E TUNSTALL
Journal:  Br Med J       Date:  1963-10-12

4.  NITROUS OXIDE AND OXYGEN FROM A SINGLE CYLINDER.

Authors:  P V COLE
Journal:  Anaesthesia       Date:  1964-01       Impact factor: 6.955

5.  Effects of cooling on the safety of premixed gases.

Authors:  J S Crawford; D B Ellis; D W Hill; J P Payne
Journal:  Br Med J       Date:  1967-04-15
  5 in total
  5 in total

1.  Resuscitation at road accidents.

Authors:  R Snook
Journal:  Br Med J       Date:  1969-11-08

2.  The use of entonox in the ambulance service.

Authors:  P J Baskett
Journal:  Proc R Soc Med       Date:  1972-01

3.  Pre-hospital analgesia with Entonox.

Authors:  N Donen; W A Tweed; D White; B Guttormson; J Enns
Journal:  Can Anaesth Soc J       Date:  1982-05

4.  Use of Entonox in the ambulance service.

Authors:  P J Baskett
Journal:  Br Med J       Date:  1970-04-04

5.  The effects of temperature on nitrous oxide and oxygen mixture homogeneity and stability.

Authors:  Patrick D Litwin
Journal:  BMC Anesthesiol       Date:  2010-10-15       Impact factor: 2.217

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.