Abstract
The distributions of within-person variances in the concentrations of 10 commonly assayed serum constituents have been derived from data on 37 healthy male subjects studied at weekly intervals over a period of five months. All 10 distributions appear to be of log-normal form. The relevance of the findings to the interpretation of differences between serial measurements in a given individual is discussed. Examples are given to show how the information on within-person variances for a particular analyte, organised into a simple graph, may be used to test medical opinions on threshold values for serial changes in the concentration of this analyte in a given individual. In this way, biological variability as well as analytical error may be taken into account quantitatively when assessing the significance of a difference between two serial measurements.Entities: