A L Barry, M B Coyle, C Thornsberry, E H Gerlach, R W Hawkinson.
Abstract
Standard Bauer-Kirby disk tests were performed with 85 selected isolates, each tested in triplicate by four different investigators. Each disk test was observed, and zone diameters were measured, under two lighting conditions (transmitted light and reflected light). The two lighting systems produced similar zone measurements (+/-2 mm) with 96% of the tests. When there were greater differences, zones appeared to be larger when observed with reflected light. Interlaboratory reproducibility was much greater when using reflected light rather than transmitted light. We concluded that zone diameters should be measured from the back of the plate while it is resting on, or held 2 to 3 inches [ca. 5.1 to 7.6 cm] above, a black, nonreflecting, flat surface, illuminated by a reflected light source.Entities:
Mesh:
Substances:
Year: 1979 PMID: 521487 PMCID: PMC273288 DOI: 10.1128/jcm.10.6.885-889.1979