Whole-blood hemoglobin determinations: a comparison of methodologies.

J A Behrens, W P Brown, D F Gibson, J C Detter.   

Abstract

The Coulter Model S provides an accurate measurement of total hemoglobin (Hb) content in peripheral blood even when elevated levels of carboxyhemoglobin (HbCO) are present. Commercial standards and diluents for spectrophotometric determinations of cyanmethemoglobin (HiCN) are variable in quality. All diluents studied produce turbidity in some samples, and this may explain why manual Hb determinations tend to be higher than hemoglobin measurement on the Coulter S. The most reliable diluent was the reagent described by Van Kampen and Zijlstra. Its reliability is enhanced by centrifugation of sample in diluent. The use of a narrow-band-pass spectrophotometer for routine HiCN determinations in a clinical laboratory is not recommended because of its sensitivity to turbidity and HbCO.

Mesh:

Substances:

Year:  1979        PMID: 517459     DOI: 10.1093/ajcp/72.6.904

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  3 in total

1.  Proficiency at hemoglobinometry in Ontario laboratories between 1975 and 1979.

Authors:  J A McBride; D E Wood; K C Carstairs; S Ezer; W I Jacobs; D C Ley; J F Murray; P H Pinkerton; A Whitney
Journal:  Can Med Assoc J       Date:  1981-07-15       Impact factor: 8.262

Review 2.  How to interpret and pursue an abnormal complete blood cell count in adults.

Authors:  Ayalew Tefferi; Curtis A Hanson; David J Inwards
Journal:  Mayo Clin Proc       Date:  2005-07       Impact factor: 7.616

3.  Evaluation of Hematological Parameters in the Genetic Prospective in Epileptic Patients of Khyber Pakhtunkhwa.

Authors:  Nazish Farooq; Niaz Ali; Shakir Ullah
Journal:  Pharmgenomics Pers Med       Date:  2019-12-23
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.